Skip to content

Newborn Ventilation in the Delivery Room: Could it be Improved With a T-piece Resuscitator?

NEWBORN VENTILATION IN THE DELIVERY ROOM: CAN IT BE IMPROVED WITH A T-PIECE RESUSCITATOR? Multicenter Cross-over Cluster Randomized Controlled Trial

Status
Completed
Phases
Phase 3
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT00443118
Enrollment
1032
Registered
2007-03-05
Start date
2009-12-31
Completion date
2012-08-31
Last updated
2018-09-28

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Resuscitation

Keywords

Bag and mask, Neopuff, Resuscitation, Newborn, Intubation, devices, PEEP

Brief summary

A multicenter cross-over cluster randomized controlled trial protocol study in newborn infants ≥ 26 weeks gestational age requiring assisted ventilation (positive pressure ventilation \[PPV\]) for resuscitation in the delivery room comparing a T-piece resuscitator device versus resuscitation bag.

Detailed description

Design: A multicenter cross-over cluster randomized controlled trial. Our hypothesis is based on the assumption that ventilating depressed newborns with a T-piece resuscitator will be more effective than SIB by increasing the proportion of resuscitated newborns with heart rate (HR) ≥ 100 beats per minute (bpm) at two minutes of life as a proxy for successful resuscitation. Population: Newborn infants ≥ 26 weeks gestational age requiring assisted ventilation (PPV) for resuscitation in the delivery room. (need for assisted ventilation at positive pressure: Heart Rate \[HR\] \< 100 beats per minute \[bpm\], apnea, gasping, cyanosis and/or hypotonia) Intervention: PPV will be performed with a T-piece resuscitator (Neopuff® group) with positive end expiratory pressure. Control: PPV will be performed with a self inflating bag (SIB group) with and without PEEP. Both devices will be used with face masks, and a peak inspiratory pressure (PIP) of 25 cm H2O will be used to begin ventilation with PEEP of 0 cm H2O in the subgroup without PEEP valve, 5 cm H2O in the subgroup with PEEP valve in the SIB group, and 5 cm H2O in the Neopuff® group. Objective: To compare the effectiveness of both instruments in reaching a heart rate of ≥ 100 bpm in depressed newborns of ≥ 26 weeks' gestational age (GA) after the initiation of positive pressure ventilation (PPV) with face mask. Primary Outcome: Proportion of newborns with HR ≥ 100 bpm at 2 minutes of life. Type of Comparison: Which of the two devices Neopuff TM or Self Inflating Bag (NP/SIB)will be more effective for ventilation of the newborn, by increasing the proportion of resuscitated newborns with heart rate (HR) ≥ 100 beats per minute (bpm) at two minutes of life as a proxy for successful resuscitation.

Interventions

DEVICET-piece resuscitator Neopuff TM

Positive pressure ventilation will be performed with Neopuff® with face mask. For this study, an initial PIP 25 cm H2O and a 5 cm H2O PEEP will be used for resuscitation according to protocol.

DEVICESelf Inflating Bag with PEEP

Positive pressure ventilation will be performed with Self Inflating Bag with PEEP with face mask. For this study, an initial PIP 25 cm H2O and a 5 cm H2O PEEP will be used for resuscitation according to protocol.

DEVICESelf Inflating Bag without PEEP

Positive pressure ventilation will be performed with Self Inflating Bag without PEEP with face mask. For this study, an initial PIP 25 cm H2O and a 5 cm H2O PEEP will be used for resuscitation according to protocol.

Sponsors

University of Alabama at Birmingham
CollaboratorOTHER
Edgardo Szyld
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
PARALLEL
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
No minimum to 2 Minutes
Healthy volunteers
No

Inclusion criteria

Live-born infants of ≥ 26 weeks' GA with HR \< 100 bpm requiring resuscitation with positive pressure ventilation with mask in the delivery room according to the current AAP/AHA recommendations

Exclusion criteria

Infants requiring intubation from birth. Newborns of \< 26 weeks´ GA. Newborns presenting major congenital malformations. Multiple births. Problems (difficulty or malfunctioning) with the device assigned.

Design outcomes

Primary

MeasureTime frame
Proportion of Infants With a HR ≥ 100 Bpm at 2 Minutes of Life.2 minutes of life

Secondary

MeasureTime frameDescription
• SpO2 Value at 2 Minutes of Life.2 minutes of life
• Proportion of Eligible Newborns Who Entered the Study and Who Were Intubated After Failure of PPV With Mask.after 2 minutes of life
• Need for Chest Compression and/or Medicationsafter 2 minutes of life
Apgar Scores at 1 and 5 Minutes1-5 minutes of lifecategorized Apgar score 1 min \<=3 and categorized Apgar score 5 min \<=5 The Apgar score is applied routinely by nurses and neonatologists to describe how vigorous the baby is at birth, it ranges from 0 to 10, with higher scores representing better outcomes.
• Incidence of Neonatal Encephalopathy During First Week of Life (Classified by Sarnat)first week of life
• Incidence of Air Leaksafter birth and during hospitalization up to four weeksincluded pneumothorax and Pneumomediastinum
Time the Newborn Takes to Reach a HR > 100 Bpm2 minutes of life
Days on Oxygenafter birth and during hospitalization up to four weeks
• Need for Mechanical Ventilation or CPAPduring hospitalization
Days on Mechanical Ventilationafter delivery and before four weeks
Days on CPAPafter birth and during hospitalization up to four weeks
• Incidence of Intracranial Hemorrhage Grades 3-4 for Preterm Newborns <32 Weeks1 day of life and 30 days after birthIncidence of intracranial hemorrhage grades 3-4 for preterm newborns \<32 weeks by ultrasound evaluation
• Use of Oxygen Treatment Beyond the Delivery Roomduring hospitalization

Countries

Argentina, Chile, Italy, Peru, United States

Participant flow

Participants by arm

ArmCount
T-Piece
Subjects assigned to be resuscitated with T-Piece
511
Self Inflating Bag With PEEP
Subjects allocated to Self Inflating Bag with PEEP
290
Self Inflating Bag Without PEEP
Subjects allocated to Self Inflating Bag without PEEP
226
Total1,027

Withdrawals & dropouts

PeriodReasonFG000FG001FG002
Overall Studyunknown primary outcome310
Overall StudyWithdrawal by Subject100

Baseline characteristics

CharacteristicT-PieceSelf Inflating Bag With PEEPSelf Inflating Bag Without PEEPTotal
Age, Continuous36 weeks
STANDARD_DEVIATION 4.1
35 weeks
STANDARD_DEVIATION 5
37 weeks
STANDARD_DEVIATION 4
36 weeks
STANDARD_DEVIATION 4.2
ANTENATAL STEROIDS
NO
369 participants206 participants154 participants729 participants
ANTENATAL STEROIDS
YES
142 participants84 participants72 participants298 participants
previous deliveries
multiparous
291 number of participants153 number of participants120 number of participants564 number of participants
previous deliveries
primiparous
220 number of participants137 number of participants106 number of participants463 number of participants
Sex: Female, Male
Female
208 Participants117 Participants101 Participants426 Participants
Sex: Female, Male
Male
303 Participants173 Participants125 Participants601 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
EG002
affected / at risk
deaths
Total, all-cause mortality
— / —— / —— / —
other
Total, other adverse events
21 / 51125 / 29019 / 226
serious
Total, serious adverse events
11 / 5118 / 2907 / 226

Outcome results

Primary

Proportion of Infants With a HR ≥ 100 Bpm at 2 Minutes of Life.

Time frame: 2 minutes of life

Population: The analysis was performed by intention to treat

ArmMeasureValue (NUMBER)
T-PieceProportion of Infants With a HR ≥ 100 Bpm at 2 Minutes of Life.93.7 percentage of participants
SIB - Self Inflating Bag With PEEP ValveProportion of Infants With a HR ≥ 100 Bpm at 2 Minutes of Life.88 percentage of participants
SIB-Without PEEP ValveProportion of Infants With a HR ≥ 100 Bpm at 2 Minutes of Life.93 percentage of participants
Secondary

Apgar Scores at 1 and 5 Minutes

categorized Apgar score 1 min \<=3 and categorized Apgar score 5 min \<=5 The Apgar score is applied routinely by nurses and neonatologists to describe how vigorous the baby is at birth, it ranges from 0 to 10, with higher scores representing better outcomes.

Time frame: 1-5 minutes of life

ArmMeasureGroupValue (NUMBER)
T-PieceApgar Scores at 1 and 5 Minutesapgar score 5 min <=530 participants
T-PieceApgar Scores at 1 and 5 Minutesapgar score 1 min <=3153 participants
SIB - Self Inflating Bag With PEEP ValveApgar Scores at 1 and 5 Minutesapgar score 5 min <=524 participants
SIB - Self Inflating Bag With PEEP ValveApgar Scores at 1 and 5 Minutesapgar score 1 min <=3111 participants
SIB-Without PEEP ValveApgar Scores at 1 and 5 Minutesapgar score 1 min <=366 participants
SIB-Without PEEP ValveApgar Scores at 1 and 5 Minutesapgar score 5 min <=523 participants
Secondary

Days on CPAP

Time frame: after birth and during hospitalization up to four weeks

ArmMeasureValue (MEAN)Dispersion
T-PieceDays on CPAP7.83 daysStandard Deviation 11.3
SIB - Self Inflating Bag With PEEP ValveDays on CPAP9.6 daysStandard Deviation 11.5
SIB-Without PEEP ValveDays on CPAP5.2 daysStandard Deviation 6.5
Secondary

Days on Mechanical Ventilation

Time frame: after delivery and before four weeks

ArmMeasureValue (MEAN)Dispersion
T-PieceDays on Mechanical Ventilation5.0 daysStandard Deviation 7.6
SIB - Self Inflating Bag With PEEP ValveDays on Mechanical Ventilation7 daysStandard Deviation 11
SIB-Without PEEP ValveDays on Mechanical Ventilation12 daysStandard Deviation 17
Secondary

Days on Oxygen

Time frame: after birth and during hospitalization up to four weeks

ArmMeasureValue (MEAN)Dispersion
T-PieceDays on Oxygen13.8 daysStandard Deviation 17
SIB - Self Inflating Bag With PEEP ValveDays on Oxygen23 daysStandard Deviation 27
SIB-Without PEEP ValveDays on Oxygen23 daysStandard Deviation 22
Secondary

• Incidence of Air Leaks

included pneumothorax and Pneumomediastinum

Time frame: after birth and during hospitalization up to four weeks

ArmMeasureValue (NUMBER)
T-Piece• Incidence of Air Leaks13 participants
SIB - Self Inflating Bag With PEEP Valve• Incidence of Air Leaks6 participants
SIB-Without PEEP Valve• Incidence of Air Leaks2 participants
Secondary

• Incidence of Intracranial Hemorrhage Grades 3-4 for Preterm Newborns <32 Weeks

Incidence of intracranial hemorrhage grades 3-4 for preterm newborns \<32 weeks by ultrasound evaluation

Time frame: 1 day of life and 30 days after birth

Population: this is an analyisis among the subgroup of subjects \<32 weeks GA. this is the population at risk of this outcome. It extremely unusual intracranial hemorrhage over 32 weeks GA.

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
T-Piece• Incidence of Intracranial Hemorrhage Grades 3-4 for Preterm Newborns <32 Weeks10 Participants
SIB - Self Inflating Bag With PEEP Valve• Incidence of Intracranial Hemorrhage Grades 3-4 for Preterm Newborns <32 Weeks1 Participants
SIB-Without PEEP Valve• Incidence of Intracranial Hemorrhage Grades 3-4 for Preterm Newborns <32 Weeks3 Participants
Secondary

• Incidence of Neonatal Encephalopathy During First Week of Life (Classified by Sarnat)

Time frame: first week of life

ArmMeasureValue (NUMBER)
T-Piece• Incidence of Neonatal Encephalopathy During First Week of Life (Classified by Sarnat)21 participants
SIB - Self Inflating Bag With PEEP Valve• Incidence of Neonatal Encephalopathy During First Week of Life (Classified by Sarnat)21 participants
SIB-Without PEEP Valve• Incidence of Neonatal Encephalopathy During First Week of Life (Classified by Sarnat)7 participants
Secondary

• Need for Chest Compression and/or Medications

Time frame: after 2 minutes of life

ArmMeasureValue (NUMBER)
T-Piece• Need for Chest Compression and/or Medications8 participants
SIB - Self Inflating Bag With PEEP Valve• Need for Chest Compression and/or Medications11 participants
SIB-Without PEEP Valve• Need for Chest Compression and/or Medications6 participants
Secondary

• Need for Mechanical Ventilation or CPAP

Time frame: during hospitalization

ArmMeasureValue (NUMBER)
T-Piece• Need for Mechanical Ventilation or CPAP116 participants
SIB - Self Inflating Bag With PEEP Valve• Need for Mechanical Ventilation or CPAP104 participants
SIB-Without PEEP Valve• Need for Mechanical Ventilation or CPAP43 participants
Secondary

• Proportion of Eligible Newborns Who Entered the Study and Who Were Intubated After Failure of PPV With Mask.

Time frame: after 2 minutes of life

ArmMeasureValue (NUMBER)
T-Piece• Proportion of Eligible Newborns Who Entered the Study and Who Were Intubated After Failure of PPV With Mask.17 percentage of participants
SIB - Self Inflating Bag With PEEP Valve• Proportion of Eligible Newborns Who Entered the Study and Who Were Intubated After Failure of PPV With Mask.28 percentage of participants
SIB-Without PEEP Valve• Proportion of Eligible Newborns Who Entered the Study and Who Were Intubated After Failure of PPV With Mask.23 percentage of participants
Secondary

• SpO2 Value at 2 Minutes of Life.

Time frame: 2 minutes of life

Population: the pulse-oximeter was reliable at 2 minutes in 69% of the cases in both groups.

ArmMeasureValue (MEAN)Dispersion
T-Piece• SpO2 Value at 2 Minutes of Life.82 percentage of oxygen saturationStandard Deviation 14
SIB - Self Inflating Bag With PEEP Valve• SpO2 Value at 2 Minutes of Life.80 percentage of oxygen saturationStandard Deviation 20
SIB-Without PEEP Valve• SpO2 Value at 2 Minutes of Life.77 percentage of oxygen saturationStandard Deviation 16
Secondary

Time the Newborn Takes to Reach a HR > 100 Bpm

Time frame: 2 minutes of life

ArmMeasureValue (MEAN)
T-PieceTime the Newborn Takes to Reach a HR > 100 Bpm1 minutes
SIB - Self Inflating Bag With PEEP ValveTime the Newborn Takes to Reach a HR > 100 Bpm1 minutes
SIB-Without PEEP ValveTime the Newborn Takes to Reach a HR > 100 Bpm1 minutes
Secondary

• Use of Oxygen Treatment Beyond the Delivery Room

Time frame: during hospitalization

ArmMeasureValue (NUMBER)
T-Piece• Use of Oxygen Treatment Beyond the Delivery Room208 participants/
SIB - Self Inflating Bag With PEEP Valve• Use of Oxygen Treatment Beyond the Delivery Room126 participants/
SIB-Without PEEP Valve• Use of Oxygen Treatment Beyond the Delivery Room96 participants/

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026