A Precision and Accuracy Study of the Codman Valve Position Verification (VPV) System.

A Precision and Accuracy Study of the Codman Valve Position Verification (VPV) System.

Status
Completed
Phases
Phase 3
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT00196196
Enrollment
274
Registered
2005-09-20
Start date
2004-09-30
Completion date
2006-10-31
Last updated
2012-05-28

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Hydrocephalus, Normal Pressure Hydrocephalus

Keywords

Hydrocephalus, Codman Hakim Programmable Valve, Programmable Valve, Shunt System

Brief summary

The purpose of this study is to determine the agreement rate between the CODMAN Valve Position Verification (VPV) adjustment outcome (reprogramming) and X-ray verification in order to support an alternative to X-ray confirmation with respect to the valve adjustment process.

Detailed description

Patients with Hydrocephalus, managed with an implanted Codman HakimTM Programmable Valve (CHPV), occasionally visit their physician to have their valve adjusted to a different setting to manage their condition. After the adjustment, an x-ray of the skull is taken to confirm the valve setting. The Codman VPV System is designed to facilitate and confirm the adjustment of a CHPV. After signing the informed consent, Subjects will be enrolled in the study and have their CHPV adjusted using the Codman VPV System. After the adjustment, an x-ray of the skull will be taken to verify the valve setting. No further follow up visit is needed after the adjustment. This study is expected to enroll about 300 Subjects from up to 20 sites in the United States.

Interventions

DEVICECodman VPV System

Valve Positioning Verification (VPV) System

Sponsors

Codman & Shurtleff
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Healthy volunteers
No

Inclusion criteria

* A CODMAN Hakim Programmable Valve (CHPV) has been implanted for the management of the Subject's Hydrocephalus. * The Subject has given written informed consent prior to enrolling in the study. * The Subject requires adjustment of the CODMAN Hakim Programmable Valve (CHPV) for the management of Hydrocephalus· * The Subject's implanted CHPV will be adjusted using the Implanted Valve Mode.

Exclusion criteria

* The Subject's valve is located in the deep lumbar position. * The Subject has an open and/or recent wound site in the region of the implanted valve. * The Subject has a history of Ultrasound gel allergies. * The Subject is a prisoner. * The subject has been previously enrolled in this protocol.

Design outcomes

Primary

MeasureTime frameDescription
Percentage of Participants Who Achieved Adjustment Complete and a Consensus X-ray ReadingDay 1Up to 5 attempts to achieve Adjustment Complete using the Codman Valve Position Verification (VPV) system for each Subject. Agreement in 2 of 3 valve X-ray readings by independent radiologists for each Subject was considered consensus X-ray. All participants included achieved Adjustment Complete and a consensus X-ray reading.

Secondary

MeasureTime frameDescription
Percentage of Agreement Between the Codman Valve Position Verification (VPV) System and Consensus X-rays Using Various ThresholdsDay 1Percentage of agreement between consensus X-ray readings and the intended setting programmed by the Codman Valve Position Verification (VPV) system. This was assessed at various steps (intervals of mmH20) as increments allow on the Codman Hakim Programmable Valve. (1 step = 10 mmH20, 2 steps = 20 mmH20, +2 steps = \>20 mmH20)

Countries

United States

Participant flow

Recruitment details

Subjects were enrolled between September 29, 2004 and October 13, 2006. Twelve (12) investigational sites within the US enrolled 274 Subjects.

Pre-assignment details

Of the 274 Subjects enrolled, 5 were determined to be ineligible - did not have hydrocephalus, a required condition per inclusion/exclusion criteria. Two (2) Subjects withdrew early. Overall, 267 Subjects were eligible for analysis.

Participants by arm

ArmCount
Codman VPV System267
Total267

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyDid not have required condition (hydroce5
Overall StudyWithdrawal by Subject2

Baseline characteristics

CharacteristicCodman VPV System
Age, Categorical
<=18 years
35 Participants
Age, Categorical
>=65 years
142 Participants
Age, Categorical
Between 18 and 65 years
90 Participants
Age Continuous57 years
STANDARD_DEVIATION 26.1
Region of Enrollment
United States
267 participants
Sex: Female, Male
Female
137 Participants
Sex: Female, Male
Male
130 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
0 / 0
serious
Total, serious adverse events
0 / 0

Outcome results

Primary

Percentage of Participants Who Achieved Adjustment Complete and a Consensus X-ray Reading

Up to 5 attempts to achieve Adjustment Complete using the Codman Valve Position Verification (VPV) system for each Subject. Agreement in 2 of 3 valve X-ray readings by independent radiologists for each Subject was considered consensus X-ray. All participants included achieved Adjustment Complete and a consensus X-ray reading.

Time frame: Day 1

ArmMeasureValue (NUMBER)
Codman VPV SystemPercentage of Participants Who Achieved Adjustment Complete and a Consensus X-ray Reading98.33 Percentage of participants
Secondary

Percentage of Agreement Between the Codman Valve Position Verification (VPV) System and Consensus X-rays Using Various Thresholds

Percentage of agreement between consensus X-ray readings and the intended setting programmed by the Codman Valve Position Verification (VPV) system. This was assessed at various steps (intervals of mmH20) as increments allow on the Codman Hakim Programmable Valve. (1 step = 10 mmH20, 2 steps = 20 mmH20, +2 steps = \>20 mmH20)

Time frame: Day 1

ArmMeasureGroupValue (NUMBER)
Codman VPV SystemPercentage of Agreement Between the Codman Valve Position Verification (VPV) System and Consensus X-rays Using Various ThresholdsPercentage of Agreement (exact match)63.89 Percentage of agreement
Codman VPV SystemPercentage of Agreement Between the Codman Valve Position Verification (VPV) System and Consensus X-rays Using Various ThresholdsPercentage of Agreement in 1 step (+/- 10 mmH20)96.11 Percentage of agreement
Codman VPV SystemPercentage of Agreement Between the Codman Valve Position Verification (VPV) System and Consensus X-rays Using Various ThresholdsPercentage of Agreement >2 steps (>20 mmH20)1.67 Percentage of agreement