Dilated Cardiomyopathy

Brief summary

This prospective study will evaluate in patients, fulfilling implant criteria for Cardiac Resynchronisation Therapy (CRT) implant, the optimal atrial contribution to the resynchronised ventricles in the event of right atrial pacing.

Detailed description

All patients will undergo advanced echocardiographic examination pre-operative, pre-discharge after implantation and at 3 and 6-months follow-up. AV-delay optimization will be performed using aortic VTI measured by continuous wave Doppler in a modified 4-chamber view. During optimisation aortic VTI will be measured at different heart rates reached by increasing atrial pacing 10, 20 and 30 beats above intrinsic heart rate (IHR).

Interventions

DEVICERenewal

Renewal is a CRT-D device for cardiac resynchronization therapy

Sponsors

Boston Scientific Corporation
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Symptoms of heart failure * Left ventricular dysfunction * Ventricular dyssynchrony * 18 years or of legal age in order to give informed consent according to national laws * Able to understand the nature of the procedure * Available for follow-up on a regular basis at an approved investigational center

Exclusion criteria

* Atrial Fibrillation * Life expectancy of less than six months due to other medical conditions * For women: pregnancy or absence of medically accepted birth control * Inability or refusal to sign the Patient Informed Consent * Inability or refusal to comply with the follow up schedule or protocol requirements * Inability to undergo device implant, including general anaesthesia if required * Mechanical tricuspid prosthesis * Currently enrolled in another investigational study, including drug investigations * Hypertrophic Obstructive Cardiomyopathy (HOCM) * Not meeting the inclusion criteria

Design outcomes

Primary

MeasureTime frameDescription
Aortic Velocity Time Integral (VTI)At implant (baseline), 3 months and 6 months Follow-upVelocity time integral of the aortic flow correlates with cardiac output and is an accepted parameter for optimization of Cardiac Resynchronization Therapy (CRT).
Optimal AV-Delay (AVD)Implant (baseline), 3 months and 6 months Follow-upOptimal AV-Delay will be measured at implant (baseline ), 3months and 6 months Follow-up for intrinsic heart rate, and 10, 20 and 30 Bpm above intrinsic heart rate.

Secondary

MeasureTime frameDescription
6 Minute Walk Testimplant (baseline), 3 months and 6 months Follow-up6 Minute Walk Test was performed at implant (baseline), 3 months and 6 months follow-up Distance walked within 6 minutes is assessed in meter. This is a test that reflects daily life activities of elderly patients.
Left Ventricular Ejection Fraction (LVEF)implant (baseline), 3 Months, 6 MonthsLeft ventricular Ejection Fraction (LVEF) was measured before implant (baseline) and at 3 and 6 months Follow-up

Countries

Denmark

Participant flow

Participants by arm

ArmCount
Group 143
Total43

Baseline characteristics

CharacteristicGroup 1
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
15 Participants
Age, Categorical
Between 18 and 65 years
28 Participants
Age, Continuous61 years
STANDARD_DEVIATION 11
Region of Enrollment
Denmark
43 participants
Sex: Female, Male
Female
12 Participants
Sex: Female, Male
Male
31 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
0 / 43
serious
Total, serious adverse events
0 / 43

Outcome results

Primary

Aortic Velocity Time Integral (VTI)

Velocity time integral of the aortic flow correlates with cardiac output and is an accepted parameter for optimization of Cardiac Resynchronization Therapy (CRT).

Time frame: At implant (baseline), 3 months and 6 months Follow-up

ArmMeasureGroupValue (MEAN)Dispersion
Cardiac Resynchronization Therapy ICD (Renewal CRT)Aortic Velocity Time Integral (VTI)VTI at baseline intrinsic rhythm22.4 cmStandard Deviation 7.7
Cardiac Resynchronization Therapy ICD (Renewal CRT)Aortic Velocity Time Integral (VTI)VTI at baseline intrinsic rhythm + 10 Bpm21.7 cmStandard Deviation 7.7
Cardiac Resynchronization Therapy ICD (Renewal CRT)Aortic Velocity Time Integral (VTI)VTI at baseline intrinsic rhythm + 20 Bpm19.4 cmStandard Deviation 7.5
Cardiac Resynchronization Therapy ICD (Renewal CRT)Aortic Velocity Time Integral (VTI)VTI at baseline intrinsic rhythm + 30 Bpm18.1 cmStandard Deviation 7.1
Cardiac Resynchronization Therapy ICD (Renewal CRT)Aortic Velocity Time Integral (VTI)VTI at 3M Follow-up intrinsic rhythm23.6 cmStandard Deviation 9
Cardiac Resynchronization Therapy ICD (Renewal CRT)Aortic Velocity Time Integral (VTI)VTI at 3M Follow-up intrinsic rhythm + 10 Bpm21.9 cmStandard Deviation 9
Cardiac Resynchronization Therapy ICD (Renewal CRT)Aortic Velocity Time Integral (VTI)VTI at 3M Follow-up intrinsic rhythm + 20 Bpm21.2 cmStandard Deviation 8.1
Cardiac Resynchronization Therapy ICD (Renewal CRT)Aortic Velocity Time Integral (VTI)VTI at 3M Follow-up intrinsic rhythm + 30 Bpm19.4 cmStandard Deviation 8.3
Cardiac Resynchronization Therapy ICD (Renewal CRT)Aortic Velocity Time Integral (VTI)VTI at 6M Follow-up intrinsic rhythm24.6 cmStandard Deviation 8.4
Cardiac Resynchronization Therapy ICD (Renewal CRT)Aortic Velocity Time Integral (VTI)VTI at 6M Follow-up intrinsic rhythm + 10 Bpm23.2 cmStandard Deviation 7.5
Cardiac Resynchronization Therapy ICD (Renewal CRT)Aortic Velocity Time Integral (VTI)VTI at 6M Follow-up intrinsic rhythm + 20 Bpm21.7 cmStandard Deviation 7
Cardiac Resynchronization Therapy ICD (Renewal CRT)Aortic Velocity Time Integral (VTI)VTI at 6M Follow-up intrinsic rhythm + 30 Bpm20.2 cmStandard Deviation 7.2
Primary

Optimal AV-Delay (AVD)

Optimal AV-Delay will be measured at implant (baseline ), 3months and 6 months Follow-up for intrinsic heart rate, and 10, 20 and 30 Bpm above intrinsic heart rate.

Time frame: Implant (baseline), 3 months and 6 months Follow-up

ArmMeasureGroupValue (MEAN)Dispersion
Cardiac Resynchronization Therapy ICD (Renewal CRT)Optimal AV-Delay (AVD)Optimal AVD at baseline intrinsic rhythm115 msStandard Deviation 24
Cardiac Resynchronization Therapy ICD (Renewal CRT)Optimal AV-Delay (AVD)Optimal AVD at baseline intrinsic rhythm +10Bpm119 msStandard Deviation 29
Cardiac Resynchronization Therapy ICD (Renewal CRT)Optimal AV-Delay (AVD)Optimal AVD at baseline intrinsic rhythm +20Bpm123 msStandard Deviation 22
Cardiac Resynchronization Therapy ICD (Renewal CRT)Optimal AV-Delay (AVD)Optimal AVD at baseline intrinsic rhythm +30Bpm127 msStandard Deviation 28
Cardiac Resynchronization Therapy ICD (Renewal CRT)Optimal AV-Delay (AVD)Optimal AVD at 3M intrinsic rhythm117 msStandard Deviation 28
Cardiac Resynchronization Therapy ICD (Renewal CRT)Optimal AV-Delay (AVD)Optimal AVD at 3M intrinsic rhythm +10Bpm117 msStandard Deviation 24
Cardiac Resynchronization Therapy ICD (Renewal CRT)Optimal AV-Delay (AVD)Optimal AVD at 3M intrinsic rhythm +20Bpm127 msStandard Deviation 28
Cardiac Resynchronization Therapy ICD (Renewal CRT)Optimal AV-Delay (AVD)Optimal AVD at 3M intrinsic rhythm +3Bpm119 msStandard Deviation 23
Cardiac Resynchronization Therapy ICD (Renewal CRT)Optimal AV-Delay (AVD)Optimal AVD at 6M intrinsic rhythm115 msStandard Deviation 24
Cardiac Resynchronization Therapy ICD (Renewal CRT)Optimal AV-Delay (AVD)Optimal AVD at 6M intrinsic rhythm +10Bpm132 msStandard Deviation 28
Cardiac Resynchronization Therapy ICD (Renewal CRT)Optimal AV-Delay (AVD)Optimal AVD at 6M intrinsic rhythm +20Bpm130 msStandard Deviation 30
Cardiac Resynchronization Therapy ICD (Renewal CRT)Optimal AV-Delay (AVD)Optimal AVD at 6M intrinsic rhythm +30Bpm130 msStandard Deviation 17
Secondary

6 Minute Walk Test

6 Minute Walk Test was performed at implant (baseline), 3 months and 6 months follow-up Distance walked within 6 minutes is assessed in meter. This is a test that reflects daily life activities of elderly patients.

Time frame: implant (baseline), 3 months and 6 months Follow-up

ArmMeasureGroupValue (MEAN)Dispersion
Cardiac Resynchronization Therapy ICD (Renewal CRT)6 Minute Walk Test6 MWT at baseline388 meterStandard Error 88
Cardiac Resynchronization Therapy ICD (Renewal CRT)6 Minute Walk Test6 MWT at 3M Follow-up470 meterStandard Error 87
Cardiac Resynchronization Therapy ICD (Renewal CRT)6 Minute Walk Test6 MWT at 6M Follow-up497 meterStandard Error 83
Secondary

Left Ventricular Ejection Fraction (LVEF)

Left ventricular Ejection Fraction (LVEF) was measured before implant (baseline) and at 3 and 6 months Follow-up

Time frame: implant (baseline), 3 Months, 6 Months

ArmMeasureGroupValue (MEAN)Dispersion
Cardiac Resynchronization Therapy ICD (Renewal CRT)Left Ventricular Ejection Fraction (LVEF)LVEF at baseline21 % of cardiac volumeStandard Deviation 6
Cardiac Resynchronization Therapy ICD (Renewal CRT)Left Ventricular Ejection Fraction (LVEF)LVEF at 3M24 % of cardiac volumeStandard Deviation 7
Cardiac Resynchronization Therapy ICD (Renewal CRT)Left Ventricular Ejection Fraction (LVEF)LVEF at 6M24 % of cardiac volumeStandard Deviation 8

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026