Neonates < 2 months' blood samples are sometimes clotted when they reach the laboratory for full blood count (FBC) analysis, leading to repeat sampling and delays in results. Pregnancy and Childbirth
Conditions
Interventions
This study looks at the feasibility of using a point-of-care testing device (Hemoscreen) for full blood count (FBC) analysis. The device will be placed on the neonatal unit, and samples will be analys
Sponsors
Harrogate and District NHS Foundation Trust
Eligibility
Sex/Gender
All
Age
0 Months to 2 Months
Inclusion criteria
Inclusion criteria: < 2 months of age. No more than 2 samples used per patient
Exclusion criteria
Exclusion criteria: Patients without consent and patients > 2 months of age
Design outcomes
Primary
| Measure | Time frame |
|---|---|
| Full blood count collected and measured at the point-of-care using the Pixcell Medical Hemoscreen device versus the hospital laboratory using a Sysmex XN1000 analyser at one time point. | — |
Secondary
| Measure | Time frame |
|---|---|
| The percentage of samples that can be processed at the point-of-care on the Pixcell Haemoscreen, but are clotted by the time of receipt in the laboratory, measured using laboratory records at one timepoint | — |
Countries
England, United Kingdom
Contacts
Public ContactNuthar Jassam
Outcome results
None listed