Perioperative hypothermia
Conditions
Interventions
Group 1: We conduct a comparative prospective observational study between three temperature measurement methods. After a patient has been included in the study, a pumonal artery catheter is placed dur
Sponsors
Universitätsklinikum Schleswig-Holstein, Campus Lübeck, Klinik für Anästhesiologie und Intensivmedizin
Eligibility
Sex/Gender
All
Age
18 Years to No maximum
Inclusion criteria
Inclusion criteria: Medical indication for the installation of a pulmonary artery catheter for perioperative monitoring
Exclusion criteria
Exclusion criteria: Impossibility of inserting a pulmonary artery catheter or contraindications to its application.
Design outcomes
Primary
| Measure | Time frame |
|---|---|
| Deviation of the temperature (in degrees Celsius) of the sensors compared to the core body temperature measurements from the pulmonary catheter. This is defined as the “limits of agreement” of the Bland-Altman analysis, which should not exceed +/- 0.5 ° C. | — |
Secondary
| Measure | Time frame |
|---|---|
| Sensitivity and specificity for hypo- and hyperthermia, the deviation of the ZHF and DS sensors from one another, as well as the measurement accuracy in different subgroups of the patient collective. | — |