Effect of different oxygen flow rates on the performance of venturi valve

EFFECT OF VARYING OXYGEN FLOW RATES ON THE DEVICE OUTPUT AND EFFECTIVE FiO2 OBTAINED AT DIFFERENT SETTINGS OF AIR ENTRAINMENT PORT OF SINGLE-UNIT VENTURI VALVE WITH MASK FOR OXYGEN THERAPY-A DESCRIPTIVE PILOT STUDY

Status
Active, not recruiting
Phases
Phase 4
Study type
Observational
Source
CTRI
Registry ID
CTRI/2017/07/009113
Enrollment
40
Registered
2017-07-26
Start date
Unknown
Completion date
Unknown
Last updated
2021-11-24

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Health Condition 1: null- post operative ASA-1 patients planned for lower limb or lower abdominal surgery under regional anaesthesia

Interventions

Intervention1: oxygen therapy by single unit venturi valve: oxygen therapy wil be given by single unit venturi valve after gentle insertion of a nasopharyngeal airway into the anaesthetised nasal cavi

Sponsors

University College of Medical Sciences and GTB Hospital Delhi
Lead Sponsor

Eligibility

Inclusion criteria

Inclusion criteria: post operative ASA-1 patients who are planned for lower limb and lower abdominal surgery under regionaL anaesthesia with normal ventilatory pattern

Exclusion criteria

Exclusion criteria: 1.smokers 2.obese patients with BMI >25 3.upper respiratory tract infection or history of uri in last 6weeks 4.deviated nasal septum/nasal polyp/any other cause of nsal obstruction 5.bleeding tendencies/past history of epistaxis/family history of bleeding disorders 6.past history of basal skull fracture.

Design outcomes

Primary

MeasureTime frame
Effective fio2 obtained by the patient with the single unit venturi valve at different settings of air entrainment port with driving oxygen flow rate of 1-15l/min in increments of 1l/minTimepoint: 30 min post-op, every 3 min thereafter for the next 2 hr

Secondary

MeasureTime frame
Concentration of oxygen output of a single unit venturi valve with driving oxygen flow rate of 1-15L/min in increments of 1L/min 2.Total gas outflow at the patient end of the single-unit venturi valve 3.Behaviour of single unit venit venturi valve in terms of its perforamance as fixed or varible at flows higher or lower than that recommended by the manufacturer.Timepoint: Before using the device on the patient, every 3 min after changing the oxygen flow in increments of 1 L/min